Moisture Analyser MA35 / Sartorius Stedim Biotech
-Fully automatic heating of a sample by two powerful metal tubular-shaped heating elements (also called dark radiators)
-Particularly simple operator guidance using self-explanatory pictograms
-Sample quantity 5-15g
-% moisture content, % dry weight , ATRO value, g residue, mg weight loss
-Shutoff criteria (optional): fully automatic, timer setting (0.1-99min.), manual
-Access to sample chamber through hinged cover with large opening angle
-1 program saved in a non-volatile memory (freely selectable parameters)
-Printout using an optional, external printer (Cat. No. 6.238 838)
-Fixed standard configuration GLP report for measuring results, calibration, adjustment (5 languages selectable)
-External calibration using optional weight sets
-Data interface RS232 C-S/V24-V28, 7 data bits (ASCII) 1 parity bit
Specifications | |
Readability: | 0.01 % or 0.001 g |
Weighing range, max: | 35 g |
Reproducibility, weight of sample 1g: Reproducibility, weight of sample 5g: | ±0.2 % ±0.05 % |
Temperature range: | 40 ... 160 °C, in increments of 1 °C |
Power consumption: | 400 VA |
Dimensions (W x D x H): | 224 x 366 x 191mm |
Max. temp.: | 160 °C |
Typemax. load
temp.
g
Resolutiong
Min.temp.
°C
PK MA35M-230NCode: 6.272 569 not in stock35
g
0.001000g
40°C
1 Code: 6.272 569 not in stock
TypeMA35M-230N
max. load
g
35g
Resolution
g
0.001000g
Min.
temp.
°C
40°C
PK1